The GENESYS 180 UV VIS spectrophotometer is a high-performance analytical instrument designed for accurate and reliable measurement of absorbance and transmittance across a wide range of wavelengths. It is widely used in research, quality control, environmental testing, pharmaceutical analysis, and academic laboratories. The instrument provides precise results for quantitative and qualitative analysis, making it suitable for various applications including concentration determination, kinetic studies, and spectral scanning.

Equipped with advanced optical components, a user-friendly interface, and a durable design, the GENESYS 180 UV VIS spectrophotometer ensures stable performance and easy operation. It features a wide wavelength range that allows flexibility for different analytical methods, while the high-resolution display and intuitive controls support quick setup and data collection. This instrument is well-suited for laboratories that require dependable and efficient UV-Vis analysis with high accuracy and repeatability.

Specifications

Baseline Flatness  ±0.002 A

Compartments

8-position cell changer (standard)


Also accommodates (optional):

4-position cell changer (long path cells)

Peltier thermostatted cell holder (20-60 °C)

Sipper accessory

Fiber optic probe coupler

ConnectionsSingle USB-A
Duplex USB-A
Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available
Print via USB, Ethernet or Wi-Fi USB adaptor available

Description

GENESYS 180 UV-Vis Spectrophotometer with UK Power Cable

Detector TypeDual Silicon Photodiodes

Display7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels

Dimensions (L x W x H)35.5 x 38.5 x 19.5 cm

Spectral Bandwidth2 nm

Wavelength Range190 to 1100 nm

Drift(At 500 nm after 1 hour warmup) <0.0005 A/hr.

Electrical RequirementsExternal AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz

For Use With (Application)Sample Concentration

KeypadTouchscreen

Lamp Life> 5 years typical, 3 years guaranteed

Noise(RMS at 500 nm 60 consecutive measurements)
≤0.00020A at 0A at 260 and 500 nm
≤0.00030A at 1A at 260 and 500 nm
≤0.00040A at 2A at 260 and 500 nm

Optical DesignDouble Beam

Photometric Accuracy±0.002 A at 0.5 A, ±0.004 A at 1.0 A, ±0.008 A at 2.0 A

Photometric Display–3A to +5A

Photometric Range–2A to +3.5A

Photometric Repeatability±0.001 A at 1 A measured at 1.0 A at 546 nm

Printer IncludedSnap-on printer available

Scan SpeedSlow, medium and fast - up to 1600 nm/min.

Stray Light< 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2)

 

Wavelength Accuracy±0.5 nm

Wavelength Data Interval0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm

Wavelength Repeatability±0.2 nm

Weight (Metric)7.5 kg

Unit SizeEach